Recall record at a glance
| Official record | Z-1932-2019 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Medtronic Navigation, Inc.-Littleton |
| Report date | Jul 10, 2019 |
| Recall initiation date | May 1, 2019 |
| Quantity | 832 units |
| Location | Littleton, MA, United States |
Official product description
Medtronic 0-arm TM 1000 Imaging Systems: Bl-700-00027-XXX and B1-700-00028-XXX (where XXX indicates the system voltage): Product Name Model#/ CFN OARM ASSY 8170000027 SYSTEM PRODUCT BI70000027 OARM ASSY BI70000027R SYS PRODUCT RWK BI70000027R BASE OARM 8170000027100 SYSTEM !DOV BI70000027100 BASE OARM Bl70000027100R SYSTEM lOOV RWK Bl70000027100R BASE OARM 8170000027120 SYSTEM 120V 8170000027120 BASE OARM Bl70000027120R SYSTEM 120V RWK BI70000027120R BASE OARM Bl70000027230 SYSTEM 230V 8170000027230 BASE OARM Bl70000027230R SYSTEM 230V RWK BI70000027230R OARM ASSY BI70000027GER SYS PRODUCT GER BI70000027GER OARM ASSY Bl70000027GERR SYS PRODUCT REFURB BI70000027GERR BASE OARM 8170000028100 SYS lOOV 8170000028100 BASE OARM BI70000028120 SYS 120V 8170000028120 BASE OARM BI70000028120R 3RD EDIT REFURB BI70000028120R BASE OARM BI70000028230 SYS 230V BI70000028230 BASE OARM BI70000028230R 3RD EDIT REFURB Bl70000028230R Product Usage: The 0-arrn Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomi~ structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Why the product was recalled
The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced
Distribution information
Worldwide Distribution - US Nationwide Foreign: ALGERIA AUSTRALIA AUSTRIA BAHRAIN BELGIUM BRAZIL BULGARIA CANADA CANADA CHILE CHINA COLOMBIA CROATIA CZECH REPUBLIC DENMARK EGYPT FINLAND FRANCE GERMANY GREECE HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KOREA KOREA, REPUBLIC OF KUWAIT LEBANON LIBYAN ARAB JAMAHIRIYA LK MEXICO NETHERLANDS NEW ZEALAND NICARAGUA POLAND PORTUGAL PUERTO RICO RUSSIAN FEDERATION SAUDI ARABIA SINGAPORE SLOVAKIA SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UNITED ARAB EMIRATES UNITED KINGDOM VENEZUELA
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.