FDA recall record

Bayer Healthcare Medrad Source Administration Sets, Catalog number INT CSS recall — FDA Z-0627-2016

The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system…

Class IIclassification
Z-0627-2016official record ID
Jan 20, 2016report date

Recall record at a glance

Official recordZ-0627-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBayer Healthcare
Report dateJan 20, 2016
Recall initiation dateNov 30, 2015
Quantity198,598
LocationIndianola, PA, United States

Official product description

Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuclear medicine diagnostic procedures.

Why the product was recalled

The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail

Distribution information

Nationwide Distribution to the states of : AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MI, MN, MO, NC, NE, NH, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT and WI.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 23, 2016

Bayer Healthcare PET Infusion Pump recall — FDA Z-0400-2017

The Source Administration Sets (SAS) used with the Medrad Intego PET Infusion System do not meet pressure testing specifications in an occlusion situation when the product is more than one…

Class IIFDA: TerminatedBayer Healthcare