Recall record at a glance
| Official record | Z-1045-2023 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | MEDLINE INDUSTRIES, LP - Northfield |
| Report date | Feb 8, 2023 |
| Recall initiation date | Dec 22, 2022 |
| Quantity | 420 units |
| Location | Northfield, IL, United States |
Official product description
MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008
Why the product was recalled
The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.
Distribution information
US Nationwide distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.