FDA recall record

MEDLINE PERFUSION SYRINGE, REF: DYNJ39223C recall — FDA Z-0886-2026

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their…

Class IIclassification
Z-0886-2026official record ID
Dec 17, 2025report date

Recall record at a glance

Official recordZ-0886-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMedline Industries, LP
Report dateDec 17, 2025
Recall initiation dateOct 27, 2025
Quantity15 units
LocationNorthfield, IL, United States

Official product description

MEDLINE PERFUSION SYRINGE, REF: DYNJ39223C

Why the product was recalled

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that certain BD Luer Tip Cap trays failed routine biocompatibility testing. Although the tip cap covers a small surface area of the syringe and does not come in direct contact with patients, there is the potential that the failure of a biocompatibility test may have varying consequences ranging from no harm or clinical impact to potential for sensitization, irritation, pyrogenicity, acute systemic toxicity or hemolysis depending on the nature and duration of contact with the syringe.

Distribution information

US Nationwide distribution in the state of CA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.