Recall record at a glance
| Official record | Z-1087-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Medline Industries, LP |
| Report date | Jan 21, 2026 |
| Recall initiation date | Dec 16, 2025 |
| Quantity | 4536 units |
| Location | Northfield, IL, United States |
Official product description
Medline medical procedure convenience kits labeled as: 1) WCH BARI-GASTRECTOMY CDS, REF CDS860237AF; 2) ROBOTIC PROCEDURE, REF CDS982128Q; 3) CUH THORACIC, REF CDS984338M; 4) THORACIC PROCEDURE, REF CDS984640I; 5) DAVINCI ROBOTIC PACK-LF, REF DYNJ0525321O; 6) LAP CHOLE PACK, REF DYNJ57640F; 7) GYN/PROSTATE ROBOTIC PACK, REF DYNJ62594D; 8) RR-DAVINCI PROSTATE ACCESSORY, REF DYNJ63392D; 9) LAPAROSCOPY PACK, REF DYNJ68187B; 10) LAPAROSCOPY PACK, REF DYNJ82368A; 11) ROBOTIC KIDNEY PACK, REF DYNJ85261; 12) DOWNTOWN ROBOTICS, REF DYNJ86178B; 13) DAVINCI LITHOTOMY PACK, REF DYNJ87610; 14) ROBOTIC WHIPPLE PACK, REF DYNJ88433; 15) PROSTATE DAVINCI, REF DYNJ900127G; 16) GENERAL LAPAROSCOPY, REF DYNJ901847M; 17) LAG TLH, REF DYNJ901890N; 18) LITHOTOMY ROBOTIC CDS, REF DYNJ903948S; 19) LAP, REF DYNJ904576B; 20) ROBOTIC, REF DYNJ905147B; 21) ROBOTICS GYN, REF DYNJ905824M; 22) WH DAVINCI TLH PROCEDURE, REF DYNJ905920B; 23) CHN OR BASIC ROBOTIC, REF DYNJ905982I; 24) KIT ROBOTIC RFD, REF DYNJ906311F; 25) ROBOTIC PROSTATE, REF DYNJ906381G; 26) DAVINCI, REF DYNJ907392C; 27) CHIASSON GASTRIC SLEEVE KIT, REF DYNJ907405C; 28) GYN, REF DYNJ908504B; 29) UROLOGY ROBOTIC, REF DYNJ908588I; 30) THORACIC ROBOTICS, REF DYNJ908777A; 31) GENERAL ROBOT, REF DYNJ908969A; 32) KIT ROBOTIC, REF DYNJ909870B; 33) ROBOTIC THORACOSCOPY, REF DYNJ910924.
Why the product was recalled
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Distribution information
US Nationwide distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.