FDA recall record

Medline Kits containing ENFIT G-TUBE CONNECTOR 1 recall — FDA Z-1691-2026

Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to…

Class IIclassification
Z-1691-2026official record ID
Apr 8, 2026report date

Recall record at a glance

Official recordZ-1691-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMedline Industries, LP
Report dateApr 8, 2026
Recall initiation dateFeb 13, 2026
Quantity9,040 kits
LocationNorthfield, IL, United States

Official product description

Medline Kits containing ENFIT G-TUBE CONNECTOR 1. SKU DYKM2720 G/J TUBE CARE KIT 2. SKU ENFITDISCKIT ENFIT DISCHARGE KIT

Why the product was recalled

Connectors were not manufactured to required dimensional specifications and may not form a proper seal with ENFit-style devices, which could lead to leakage. Leakage may result in delayed care and fluid loss. Additional consequences may also occur if user over-tightens the connector to stop the leakage, including additional clinical intervention, risk of infection, patient discomfort, and prolonged recovery process.

Distribution information

Worldwide - US Nationwide and the countries of Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.