Recall record at a glance
| Official record | Z-1811-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Ongoing |
| Recalling firm | MEDLINE INDUSTRIES, LP - Northfield |
| Report date | Jun 11, 2025 |
| Recall initiation date | Apr 22, 2025 |
| Quantity | 1,853 total |
| Location | Northfield, IL, United States |
Official product description
Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ACCESSORY PACK SKU DYNJ59270 2. CRANI PACK-LF SKU DYNJ44805M 3. CRANIOTOMY SKU CDS982719V CDS982719W DYNJ904168F DYNJ908404B DYNJ908723A DYNJ908723C 4. CRANIOTOMY #IMF 56646-LF SKU DYNJVB91001A 5. CRANIOTOMY CDS SKU CDS981888X CDS983467I 6. CRANIOTOMY PACK-LF SKU DYNJ09882G DYNJ09882I 7. CRANIOTOMY S NASSAU SKU DYNJ905879G 8. MHC CRANIOTOMY SKU DYNJ903713K DYNJ903713L 9. PAD CRANIOTOMY SKU DYNJ902149J
Why the product was recalled
Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.
Distribution information
US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.