FDA recall record

Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item recall — FDA Z-0765-2025

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber…

Class Iclassification
Z-0765-2025official record ID
Jan 22, 2025report date

Recall record at a glance

Official recordZ-0765-2025
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmMEDLINE INDUSTRIES, LP - Northfield
Report dateJan 22, 2025
Recall initiation dateDec 16, 2024
Quantity3930 units
LocationNorthfield, IL, United States

Official product description

Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135,CDS,IV,TUBE-JEFFERSON UNIV., Item Number: 60232481; 9) MTS,135,YELLOW,MICRO,SPIKE,LUTHERAN MED, Item Number: 60120583; 10) MTS,ANGIO,NEURO,KIT,NORTH SHORE UNIV, Item Number: 601403920; 11) MTS,CUST. FILTER-NEWYORK UNIVERSITY, Item Number: 62140602; 12) MTS,INTERV,NEURO,KIT-WESTCHESTER,MC, Item Number: 602301614; 13) MTS,NEU,NORTH SHORE HOSPITAL, Item Number: 6010771601; 14) MTS,NEURO,KIT, Item Number: 600602914; 15) MTS,NEURO,KIT,2,NEWENGLAND MED CTR, Item Number: 601400111; 16) MTS,NEURO,KIT,YLW&RED,FLORIDA HOSP, Item Number: 600604423; 17) MTS,NEURO,KIT-JOHN,F,KENNEDY MED CENTER, Item Number: 60110552; 18) MTS,NEURO,KIT-OVERLOOK HOSPITAL PG, Item Number: 60150035; 19) MTS,NEURO,SPIKE,KIT,CAPITAL HLTHSYS-FULD, Item Number: 60182723; 20) MTS,NEURO-RADIOLOGY,KIT-YALE-NEWHAVEN, Item Number: 602500214; 21) MTS,STROKE/EMBO,KIT,NORTH SHORE UNIV, Item Number: 601403921; 22) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601305210; 23) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601400110;

Why the product was recalled

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

Distribution information

US Nationwide Distribution

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records