FDA recall record

Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS recall — FDA Z-1397-2026

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the…

Class IIclassification
Z-1397-2026official record ID
Feb 25, 2026report date

Recall record at a glance

Official recordZ-1397-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMedline Industries, LP
Report dateFeb 25, 2026
Recall initiation dateJan 7, 2026
Quantity4757 units
LocationNorthfield, IL, United States

Official product description

Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 3) KIT PLASTIC INSERT TISSUE EXPA, Model Number: DYKMBNDL158; 4) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170; 5) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170A; 6) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170B; 7) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182; 8) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182A; 9) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182B; 10) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183; 11) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183A; 12) KIT PLASTIC ROTATIONAL FLAP, Model Number: DYKMBNDL183B; 13) KIT STC FACIAL FRACTURE, Model Number: DYKMBNDL34A; 14) KIT SURG ONC MASTECTOMY A, Model Number: DYKMBNDL39; 15) KIT SURG ONC MASTECTOMY A, Model Number: DYKMBNDL39A; 16) KIT SURG ONC MASTECTOMY, Model Number: DYKMBNDL39C; 17) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL46; 18) KIT STC PLASTIC FREE FLAP, Model Number: DYKMBNDL54; 19) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69; 20) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69A; 21) KIT PLASTICS BREAST REDUCTION, Model Number: DYKMBNDL69B; 22) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78; 23) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78A; 24) KIT PLASTICS BREAST AUGMENTATI, Model Number: DYKMBNDL78B; 25) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84; 26) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84C; 27) MJR PLASTICS PK MV SURG CT-LF, Model Number: DYNJ0785310G; 28) ASC ENT PLASTIC PACK-LF, Model Number: DYNJ0854650J; 29) BREAST PACK ASC-LF, Model Number: DYNJ36469I; 30) ENT TYMPANOPLASTY PACK-LF, Model Number: DYNJ38988J; 31) DR WILDE BRSTMASS/PARTMAST-XM, Model Number: DYNJ39831F; 32) BREAST RECONSTRUCTION-HOULE, Model Number: DYNJ44852K; 33) BREAST RECONSTRUCTION-HOULE, Model Number: DYNJ44852M; 34) BREAST PACK, Model Number: DYNJ50957G; 35) PLASTICS BREAST PACK, Model Number: DYNJ55358G; 36) BHN IMPLANT PACK, Model Number: DYNJ56737A; 37) BREAST PACK, Model Number: DYNJ60210A; 38) PLASTICS PACK 2, Model Number: DYNJ64927F; 39) SPECIALTY FACIAL PLASTICS PACK, Model Number: DYNJ68844A; 40) BREAST IMPLANT PACK, Model Number: DYNJ69351; 41) BREAST/TUMMY PACK, Model Number: DYNJ83788; 42) HMSL PLASTICS, Model Number: DYNJ83884B; 43) CHEST/BREAST-LF, Model Number: DYNJ83950A; 44) CHEST/BREAST-LF, Model Number: DYNJ83950B; 45) FREE FLAP A DIRTY, Model Number: DYNJ86165; 46) MAJOR SURGERY PACK, Model Number: DYNJ86816; 47) BREAST REDUCTION PACK, Model Number: DYNJ87444; 48) OMF FREE FLAP, Model Number: DYNJ902005I; 49) BREAST CDS, Model Number: DYNJ902558G; 50) TRUNCAL CONTOURING, Model Number: DYNJ903555D; 51) TRUNCAL CONTOURING, Model Number: DYNJ903555F; 52) NEW TISSUE RETRIEVAL MH, Model Number: DYNJ903849C; 53) LARGE PROCEDURE-LF, Model Number: DYNJ907847A; 54) (34) STC FACIAL FRACTURE, Model Number: DYNJ907874A; 55) FACELIFT, Model Number: DYNJ910462; 56) MAJOR PLASTIC BREAST PACK-LF, Model Number: DYNJT3283; 57) MAJOR BREAST PACK-LF, Model Number: DYNJT3417; 58) PLASTICS PACK-LF, Model Number: PHS806973G

Why the product was recalled

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Distribution information

Worldwide distribution - US Nationwide.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.