Medline Convenience Kits: 1) OPEN HEART CDS recall — FDA Z-1390-2026

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the…

Class IIclassification
Z-1390-2026official record ID
Feb 25, 2026report date

Recall record at a glance

Official recordZ-1390-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMedline Industries, LP
Report dateFeb 25, 2026
Recall initiation dateJan 7, 2026
Quantity39315 units
LocationNorthfield, IL, United States

Official product description

Medline Convenience Kits: 1) OPEN HEART CDS, Model Number: CDS840015AQ; 2) OPEN HEART CDS, Model Number: CDS840023T; 3) OFF PUMP CABG CDS, Model Number: CDS840087AI; 4) OFF PUMP CABG CDS, Model Number: CDS840087AK; 5) OFF PUMP CABG CDS, Model Number: CDS840087AL; 6) OPEN HEART, Model Number: CDS840150S; 7) OPEN HEART, Model Number: CDS840150T; 8) OPEN HEART, Model Number: CDS840150U; 9) OPEN HEART, Model Number: CDS840150V; 10) OPEN HEART, Model Number: CDS840150W; 11) UNIVERSAL HEART CDS, Model Number: CDS840246O; 12) GENERAL AAA #11-RF, Model Number: CDS840261AB; 13) OPEN HEART ADULT, Model Number: CDS840396AA; 14) OPEN HEART ADULT, Model Number: CDS840396AB; 15) OPEN HEART ADULT, Model Number: CDS840396X; 16) OPEN HEART ADULT, Model Number: CDS840396Y; 17) OPEN HEART CDS, Model Number: CDS840428K; 18) OPEN HEART CDS, Model Number: CDS840428L; 19) PACEMAKER CDS, Model Number: CDS980839C; 20) KIT, CVC ADD A PACK HUM-LF, Model Number: CDS982564C; 21) OPEN HEART ADULT HUM, Model Number: CDS982566K; 22) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575K; 23) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575M; 24) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575O; 25) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575P; 26) OPEN HEART, Model Number: CDS983348S; 27) HEART CABG CDS, Model Number: CDS983376Q; 28) HEART CABG CDS, Model Number: CDS983376T; 29) HEART CABG CDS, Model Number: CDS983376U; 30) HEART CABG CDS, Model Number: CDS983376V; 31) OPEN HEART CDS, Model Number: CDS983513F; 32) OPEN HEART ADULT PART 2 CDS, Model Number: CDS983786D; 33) OPEN HEART ADULT PART 2 CDS, Model Number: CDS983786F; 34) OPEN HEART ADULT PART 2 CDS, Model Number: CDS983786G; 35) PEDS CARDIAC CDS, Model Number: CDS984117F; 36) OPEN HEART CDS, Model Number: CDS984276P; 37) OPEN HEART ACCESSORY CDS, Model Number: CDS984289J; 38) OPEN HEART ACCESSORY CDS, Model Number: CDS984289K; 39) OPEN HEART ACCESSORY CDS, Model Number: CDS984289L; 40) OPEN HEART ACCESSORY CDS, Model Number: CDS984289M; 41) OPEN HEART ACCESSORY CDS, Model Number: CDS984289N; 42) OPEN HEART ACCESSORY CDS, Model Number: CDS984289O; 43) OPEN HEART CDS, Model Number: CDS984355F; 44) OPEN HEART CDS, Model Number: CDS984355G; 45) OPEN HEART CDS, Model Number: CDS984355I; 46) OPEN HEART CDS, Model Number: CDS984355J; 47) OPEN HEART A, Model Number: CDS985105M; 48) OPEN HEART A, Model Number: CDS985105N; 49) OPEN HEART A, Model Number: CDS985105O; 50) VASCULAR CDS, Model Number: CDS985215C; 51) BASE HEART PACK-LF, Model Number: CVT159AA; 52) ENDO KIT, Model Number: DYKE1580B; 53) KIT VASCULAR MINOR, Model Number: DYKMBNDL11; 54) KIT VASCULAR CAROTID, Model Number: DYKMBNDL118; 55) KIT VASCULAR CAROTID, Model Number: DYKMBNDL118A; 56) KIT VASCULAR BYPASS GRAFT GENE, Model Number: DYKMBNDL120; 57) KIT VASCULAR PERCUTANEOUS TEVA, Model Number: DYKMBNDL14; 58) KIT VASCULAR PERCUTANEOUS TEVA, Model Number: DYKMBNDL14A; 59) KIT VASCULAR VNS, Model Number: DYKMBNDL152; 60) KIT VASCULAR VNS, Model Number: DYKMBNDL152A; 61) KIT THORACIC SYMPHATECTOMY, Model Number: DYKMBNDL187; 62) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189B; 63) KIT VASCULAR ENDOVENOUS ABLATI, Model Number: DYKMBNDL191; 64) KIT VASCULAR ANGIOGRAM GENERIC, Model Number: DYKMBNDL192; 65) KIT VASCULAR ENDOVENOUS ABLATI, Model Number: DYKMBNDL203; 66) KIT VASCULAR MEDIAN ARCUATE, Model Number: DYKMBNDL209A; 67) KIT VASCULAR MEDIAN ARCUATE, Model Number: DYKMBNDL209B; 68) CARDIAC ADULT CABG, Model Number: DYKMBNDL20A; 69) CARDIAC ADULT CABG, Model Number: DYKMBNDL20B; 70) KIT CARDIAC CSICU ADULT ECMO, Model Number: DYKMBNDL21; 71) KIT VASCULAR AAA REPAIR, Model Number: DYKMBNDL23A; 72) KIT CARDIAC ANEURYSM, Model Number: DYKMBNDL27A; 73) KIT CARDIAC ANEURYSM, Model Number: DYKMBNDL27B; 74) KIT PEDS CARDIAC ADULT, Model Number: DYKMBNDL4

Why the product was recalled

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Distribution information

Worldwide distribution - US Nationwide.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.