Recall record at a glance
| Official record | Z-0723-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | MEDLINE INDUSTRIES, LP - Northfield |
| Report date | Jan 1, 2025 |
| Recall initiation date | Nov 6, 2024 |
| Location | Northfield, IL, United States |
Official product description
Medline brand, medical procedure convenience kits, labeled as: 1) CRANIOTOMY, SKU CDS982719X; 2) MVFF/FIBU #54-RF, SKU DYNJ21927Q; 3) MVFF/FIBU #54-RF, SKU DYNJ21927S; 3) NEURO PEDIATRIC, SKU DYNJ61607A; 4) CRANIOTOMY PACK, SKU DYNJ66419C; 5) LYMPHATIC BYPASS PACK, SKU DYNJ69907; 6) LYMPHATIC BYPASS PACK, SKU DYNJ69907A; 7) HEAD AND NECK PACK - 51597, SKU DYNJ81934B; 8) MAJOR CRANIOTOMY PACK, SKU DYNJ82007A; 9) SPINE PACK, SKU DYNJ82011A; 10) CRANIOTOMY MERCY STK 256, SKU DYNJ84068; 11) WMC UNIVERSAL NEURO PACK, SKU DYNJ84496A; 12) SPINE PACK LF, SKU DYNJ85830; 13) AHT SPINE PACK, SKU DYNJ903566D; 14) LAMINECTOMY SPINAL FUSION PACK, SKU DYNJ909364A; 15) DIEP, SKU DYNJ909472; 16) AMBULATORY SINUS, SKU DYNJ910292; 17) AMBULATORY SINUS, SKU DYNJ910292A
Why the product was recalled
Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods
Distribution information
US Nationwide. Canada.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.