FDA recall record

American Contract Systems Medical convenience kits Item number/ Product description AKGN82E recall — FDA Z-0172-2026

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection…

Class IIclassification
Z-0172-2026official record ID
Oct 22, 2025report date

Recall record at a glance

Official recordZ-0172-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmAmerican Contract Systems Inc.
Report dateOct 22, 2025
Recall initiation dateSep 4, 2025
Quantity32,433 total
LocationGrimes, IA, United States

Official product description

Medical convenience kits Item number/ Product description AKGN82E GENERAL AMCB08Y CHEST BREAST PACK ANBP10AC BASIC PACK - 205939 ANIR77AM INTERVENTIONAL RADIOLOGY PACK - 205953 CFLA39G LAPAROTOMY PACK CIBL33I BASIC LAP PACK EVMN24E MINOR ESC EVST48E STARTER FDTA11AA T & A PACK FHHY25P HYSTEROSCOPY PACK HGML99G MAJOR LAP PACK HGXR41H X-RAY PACK HISU54M SET-UP PACK IHAA19AP PK A A A IHMP89K MINOR PROCEDURE PACK IHPD14AM PK PEDIATRIC LKVG10O VAG DELIVERY LMAG20G VASCULAR PACK LMLE46AA EXTREMITY LOWER PK MIEN58A DTC GENERAL ENDOSCOPY PACK(PS 209468) NUMB04H MAJOR BASIN SET 314666 OWBS11U BASIN SET - 206021 RCMS71B MIS SPECIALTY RCH SFMB55X MINOR BASIN - 208293 STBA09F BASIN SET TNBS21O BASIN SET 612833 UDBH37AX BASIC HEART PACK - 206044 UDMA50X MINOR ADULT - 206067 UICR54W ACS CRS PACK (PS055042) UIPC89D PROCUREMENT PACK UIPT13F PITCHER (PS 907953) UISB96AB SMALL BACK TABLE SETUP (PS 907419)

Why the product was recalled

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Distribution information

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.