FDA recall record

American Contract Systems Medical convenience kits Item Number/Description ANSH14AF SHOULDER recall — FDA Z-0184-2026

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection…

Class IIclassification
Z-0184-2026official record ID
Oct 22, 2025report date

Recall record at a glance

Official recordZ-0184-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmAmerican Contract Systems Inc.
Report dateOct 22, 2025
Recall initiation dateSep 4, 2025
Quantity32,433 total
LocationGrimes, IA, United States

Official product description

Medical convenience kits Item Number/Description ANSH14AF SHOULDER PACK - 205962 EVAH76H ANTERIOR HIP PACK FDSA15M SHOULDER ARTHROSCOPY GRTS90G TOTAL SHPULDER PACK HGHI26G HIP PINNING PACK HJTO25C TOTAL HIP ACCESSORY PACK HNEX60L EXTREMITY PACK IAHD53C HAND PACK IASH65A SHOULDER PACK IHUE40AI UPPER EXTREMITY PK| STRL F G ILTJ47AA TOTAL JOINT PACK LMHP14Z ORIF HIP, ST LUKES METHODIS LMSN44AE SPINAL PACK MHHP25H HAND PACK - 206002 MHTH90AJ TOTAL HIP PACK - 206016 MWAH32P ANTERIOR HIP PACK MWTK13AH TOTAL KNEE PLOR89H ORTHO MINOR PACK RCUE81B UPPER EXTREMITY RCH 0247031 TSTH54D TOTAL HIP TRAY UDEX85Y DSC EXTREMITY - 206050 UDHD69U HAND PACK - 206055 UDKN68AH CUSTOM KNEE - 206061 UIHN18AW ACS HAND PACK (PS 020640) UIHN18AX ACS HAND PACK (PS 020640) UIKN28AR ASC KNEE PACK (PS 020641) UISO44I SFCH SMALL ORTHO PACK (PS 143213)

Why the product was recalled

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Distribution information

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.