FDA recall record

Medica Capillary Tubes recall — FDA Z-1370-2026

A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when…

Class IIclassification
Z-1370-2026official record ID
Feb 25, 2026report date

Recall record at a glance

Official recordZ-1370-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMedica Corporation
Report dateFeb 25, 2026
Recall initiation dateDec 31, 2025
Quantity1,578 units
LocationBedford, MA, United States

Official product description

Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.

Why the product was recalled

A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.

Distribution information

International distribution to the countries of Russia, Bulgaria, Kazakhstan, Ukraine, Pakistan, Belarus.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.