FDA recall record

Owens & Minor Distribution Medi Choice Arm Slings ASL 2103 Small 7.5 x 12.5 Inch 1/EA, 24 recall — FDA Z-0888-2017

Owens & Minor is the initial importer and private label distributor of Medi Choice Arm Slings (Product). The Product is manufactured in China by Tops…

Class IIclassification
Z-0888-2017official record ID
Dec 28, 2016report date

Recall record at a glance

Official recordZ-0888-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmOwens & Minor Distribution, Inc.
Report dateDec 28, 2016
Recall initiation dateOct 31, 2016
Quantity568 cases
LocationMechanicsville, VA, United States

Official product description

Medi Choice Arm Slings ASL 2103 Small 7.5 x 12.5 Inch 1/EA, 24 ENCA ASL 2104 Medium 8 x 15 Inch 1/EA, 24 ENCA ASL 2105 Large 8.5 x 17.5 Inch 1/EA, 24 ENCA ASL 2106 Extra-Large 9 x 20 Inch 1/EA, 24 ENCA ASL 2150 Universal 7 x 18 Inch 1/EA, 24 ENCA Medical - An arm sling is a device intended for medical purposes to immobilize the arm, by means of a fabric band suspended from around the neck.

Why the product was recalled

Owens & Minor is the initial importer and private label distributor of Medi Choice Arm Slings (Product). The Product is manufactured in China by Tops Medical Articles Nantong Inc. Owens & Minor first learned of the product issue that prompted the product removal action via customer complaints received in September 2016. Two customer complaints received on September 1"1, 2016 reported surface contamination on the arm slings. Two additional complaints reported on September 19, 2016 and October 7, 2016 reported the same issue. Upon receipt of the complaints on September 1, 2016, Owens & Minor immediately initiated a CAPA investigation for root cause analysis and determination of corrective/preventive actions. A quality inspection of inventory confirmed that the surface contamination was confined to lots manufactured in May, June, and July of 2016. The surface contamination for these lots was found to be frequent. Subsequent August product lots were inspected and no surface contamination was found. Limited inventory from lots prior to May 2016 were available and inspected as well. No surface contamination was found on these prior lots. Owens & Minor has quarantined the affected product lots. The product is on hold at all Owens & Minor Distribution Centers (DC). Samples of contaminated product have been sent to the manufacturer for evaluation and to a certified testing lab for identification of contaminate. These results are pending at this time. Based on the frequency of the product defect as well as evaluation of potential adverse health event risk, Owens & Minor initiated a voluntary recall for the above stated affected product lots.

Distribution information

US Nationwide Distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Telephone.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.