FDA recall record

Smiths Medical ASD Medfusion Syringe Pump, Model 3500-500, v3, v4, v5, and v6 recall — FDA Z-0076-2024

A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System…

Class IIclassification
Z-0076-2024official record ID
Oct 18, 2023report date

Recall record at a glance

Official recordZ-0076-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmSmiths Medical ASD Inc.
Report dateOct 18, 2023
Recall initiation dateAug 16, 2023
Quantity20815 pumps
LocationMinneapolis, MN, United States

Official product description

Medfusion Syringe Pump, Model 3500-500, v3, v4, v5, and v6

Why the product was recalled

A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Failure Alarm (including Force Sensor BGND Test, Force Sensor Bridge Test, or Force Sensor Test). However, if the calibration shift is not large enough to trigger a System Failure Alarm, the threshold to detect an occlusion may increase, increasing the time to occlusion detection, or the threshold may decrease, leading to false occlusion alarms. Although shifts in the force sensor calibration may occur over time with any device, an increased potential for such shifts has been reported in devices produced before April 2022 due to mechanical interference between parts of the plunger head assembly. Out of an abundance of caution, we are notifying all customers of this potential issue.

Distribution information

Worldwide distribution - US Nationwide and the country of Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.