FDA recall record

CareFusion 303 MEDFLEX, REF: 1119-00 recall — FDA Z-1552-2025

If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first…

Class IIclassification
Z-1552-2025official record ID
Apr 23, 2025report date

Recall record at a glance

Official recordZ-1552-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmCareFusion 303, Inc.
Report dateApr 23, 2025
Recall initiation dateOct 5, 2023
Quantity2679
LocationSan Diego, CA, United States

Official product description

MEDFLEX, REF: 1119-00; MEDBANK MAIN, REF: 1137-00; MEDFLEX 2.0, REF: 1139-00; MEDBANK TOWER MAIN, REF: 1145-00; MEDBANK MINI MAIN, REF: 1147-00 BD PYXIS MEDBANK Facility Software, REF: 139090-01; BD PYXS MEDBNK TWR MN 2HH1FH2HM2FM2FG1FF, REF: 169-100; BD PYXIS MEDBANK MINI 1HH-1FH- 1HM, REF: 169-101; BD PYXIS MEDBANK MINI 1HH-1HM- 1FM, REF: 169-102; BD PYXIS MEDBANK MINI 1HH-3HM, REF: 169-103; BD PYXIS MEDBANK MINI 2FH, REF: 169-104; BD PYXIS MEDBANK MINI 2FM, REF: 169-105; BD PYXIS MEDBANK MINI 2HH-1FH, REF: 169-106; BD PYXIS MEDBANK MINI 2HH-1FM , REF: 169-107; BD PYXIS MEDBANK MINI 2HH-2HM, REF: 169-108; BD PYXIS MEDBANK MINI 2HM-1FM, REF: 169-109; BD PYXIS MEDBANK MINI 3HH-1FG, REF: 169-110; BD PYXIS MEDBANK MINI 3HH-1HM, REF: 169-111; BD PYXIS MEDBANK MINI 4HH, REF: 169-112; BD PYXIS MEDBANK MINI 4HM, REF: 169-113; BD PYXIS MDBANK TWR MN 7HH-1HM-3FM RxNow, REF: 169-132; BD PYXIS MEDBANK MINI 1FH-1FM, REF: 169-137; BD PYXIS MEDBANK TWR MN 1FH- 10HM-1FM, REF: 169-138; BD PYXIS MEDBANK TWR MN 2HH- 4FH-2FM, REF: 169-139; OBS BD PYXIS MB TOWER MAIN, REF: 169-153; OBS BD PYXIS MB TWR RXNOW MAIN W/MED PRI, REF: 169-155; OBS BD PYXIS MB TWR RXNOW MAIN W/PAT PRI, REF: 169-156; OBS BD PYXIS MB MINI MAIN W/CUBIE DRWS, REF: 169-159; OBS BD PYXIS MB MINI MAIN W/MATRIX DRWS, REF: 169-160; OBS BD PYXIS MB MINI MAIN W/HH FALSE DRW, REF: 169-161; OBS BD PYXIS MB MINI MAIN W/FH FALSE DRW, REF: 169-162; BD PYXIS MEDBANK TWR MN 10HH- 1FH-1FM, REF: 169-23; BD PYXIS MEDBANK TWR MN 10HH- 2FH, REF: 169-24; BD PYXIS MEDBANK TWR MN 10HH- 2FM, REF: 169-25; BD PYXIS MEDBANK TWR MN 12HH- 1FF, REF: 169-26; BD PYXIS MEDBANK TWR MN 12HH- 1FH, REF: 169-27; BD PYXIS MEDBANK TWR MN 12HH- 1FM, REF: 169-28; BD PYXIS MEDBANK TWR MN 14HH, REF: 169-29; BD PYXIS MEDBANK TWR MN 1HH- 1HM-2FM-4FF, REF: 169-30; BD PYXIS MEDBANK TWR MN 2FH- 2HM-4FM, REF: 169-31; BD PYXIS MEDBANK TWR MN 2HH- 10HM-1FM, REF: 169-32; BD PYXIS MEDBANK TWR MN 2HH- 1FH-5FM, REF: 169-33; BD PYXIS MEDBANK TWR MN 2HH- 2FH-4FM, REF: 169-34; BD PYXIS MEDBANK TWR MN 2HH- 2FM-4FF, REF: 169-35; BD PYXIS MEDBANK TWR MN 2HH- 4HM-4FF, REF: 169-36; BD PYXIS MEDBANK TWR MN 2HH- 4HM-4FM, REF: 169-37; BD PYXIS MEDBANK TWR MN 2HH- 6FM, REF: 169-38; BD PYXIS MEDBANK TWR MN 2HH- 6HM-3FM, REF: 169-39; BD PYXIS MEDBANK TWR MN 2HH- 8HM-2FM, REF: 169-40; BD PYXIS MEDBANK TWR MN 2HM- 6FM, REF: 169-41; BD PYXIS MEDBANK TWR MN 3FH-4FM , REF: 169-42; BD PYXIS MEDBANK TWR MN 3HH- 7HM-2FM, REF: 169-43; BD PYXIS MEDBANK TWR MN 4HH- 10HM, REF: 169-44; BD PYXIS MEDBANK TWR MN 4HH- 1FH-2HM-3FM, REF: 169-45; BD PYXIS MEDBANK TWR MN 4HH- 1FH-4FM, REF: 169-46; BD PYXIS MEDBANK TWR MN 4HH- 1FH-4HM-2FM, REF: 169-47; BD PYXIS MEDBANK TWR MN 4HH- 1FH-6HM-1FM, REF: 169-48; BD PYXIS MEDBANK TWR MN 4HH- 2FH-2HM-2FM, REF: 169-49; BD PYXIS MEDBANK TWR MN 4HH- 2FH-3FM, REF: 169-50; BD PYXIS MEDBANK TWR MN 4HH- 2HM-4FM, REF: 169-51; BD PYXIS MEDBANK TWR MN 4HH- 3FH-2FM, REF: 169-52; BD PYXIS MEDBANK TWR MN 4HH- 4FH-1FM, REF: 169-53; BD PYXIS MEDBANK TWR MN 4HH- 4HM-3FM, REF: 169-54; BD PYXIS MEDBANK TWR MN 4HH-5FH, REF: 169-55; BD PYXIS MEDBANK TWR MN 4HH- 5FM, REF: 169-56; BD PYXIS MEDBANK TWR MN 4HH- 6HM-2FM, REF: 169-57; BD PYXIS MEDBANK TWR MN 4HH- 8HM-1FM, REF: 169-58; BD PYXIS MEDBANK TWR MN 4HM- 5FM, REF: 169-59; BD PYXIS MEDBANK TWR MN 5HH- 3HM-3FM, REF: 169-60; BD PYXIS MEDBANK TWR MN 5HH- 9HM, REF: 169-61; BD PYXIS MEDBANK TWR MN 6FH-1FM, REF: 169-62; BD PYXIS MEDBANK TWR MN 6HH- 1FH-2HM-2FM, REF: 169-63; BD PYXIS MEDBANK TWR MN 6HH- 1FH-3FM, REF: 169-64; BD PYXIS MEDBANK TWR MN 6HH- 2FH-2FM, REF: 169-65; BD PYXIS MEDBANK TWR MN 6HH- 2FH-2HM-1FM, REF: 169-66; BD PYXIS MEDBANK TWR MN 6HH-4FH, REF: 169-67; BD PYXIS MEDBANK TWR MN 6HH- 4FM, REF: 169-68; BD PYXIS MEDBANK TWR MN 6HH- 4HM-2FM, REF: 169-69; BD PYXIS MEDBANK TWR MN 6HH- 6HM-1FM, REF: 169-70; BD PYXIS MEDBANK TWR MN 6HH- 8HM, REF: 169-71; BD PYXIS MEDBANK TWR MN 7FH, REF: 169-72; BD PYXIS M

Why the product was recalled

If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.

Distribution information

US Nationwide distribution in the states of KS, TX, CA, PA, NJ, FL, IN, AZ, IL, MA, NY, KY, UT, MD, WA, OH, AL.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.