Recall record at a glance
| Official record | Z-0095-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Advance-Esthetic LLC |
| Report date | Oct 18, 2023 |
| Recall initiation date | Oct 9, 2023 |
| Quantity | 19 |
| Location | Fort Lauderdale, FL, United States |
Official product description
MED-810A Zemits NDPrime Laser products
Why the product was recalled
Advance-Esthetic LLC failed to comply with the applicable standards regarding Electronic Product Radiation Control (EPRC) regulations associated with the laser product performance standards and certification. This Recall provides correction by the firm regarding Zemits NdPrime Yag Tattoo Removal Laser. The product requires update or is missing the following in part: remote interlock connector, safety interlocks, manual reset mechanism, beam attenuator, tests for determination of compliance, certification, class IV designation and warning, labels, and labeling.
Distribution information
US Nationwide Distribution
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: FDA Mandated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.