McKesson Cardiology Hemo recall — FDA Z-0093-2015

McKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and…

Class IIclassification
Z-0093-2015official record ID
Oct 29, 2014report date

Recall record at a glance

Official recordZ-0093-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMckesson Information Solutions LLC
Report dateOct 29, 2014
Recall initiation dateJun 25, 2014
Quantity625 devices
LocationAlpharetta, GA, United States

Official product description

McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.

Why the product was recalled

McKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and to identify in a timely manner whether a parameter box is out of calibration. McKesson has identified that its communication to customers regarding the required maintenance tests was insufficient. While each unit has an attached sticker i

Distribution information

Worldwide Distribution-USA (nationwide) and the countries of Australia, Botswana, Canada, Germany, New Zealand, Netherlands, Pakistan, Portugal, South Africa, Turkey and The United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.