Recall record at a glance
| Official record | Z-2359-2019 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Agilent Technologies, Inc. |
| Report date | Sep 4, 2019 |
| Recall initiation date | Aug 15, 2018 |
| Quantity | 235 |
| Location | Santa Clara, CA, United States |
Official product description
MassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometers, models K6460 and K6420 Product Usage: A mass spectrometer for clinical use is a device intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass.
Why the product was recalled
When utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made.
Distribution information
US Nationwide. distribution.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.