FDA recall record

Masimo SET uSpO2 Cable The Masimo SET uSpO2 Pulse Oximetry Cable is an in-line cable that recall — FDA Z-0213-2015

Firm has identified a small number of Oximetry Cables with crossed-internal wires. The effect on performance when these wires are crossed is that when a…

Class IIclassification
Z-0213-2015official record ID
Nov 19, 2014report date

Recall record at a glance

Official recordZ-0213-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMasimo Corporation
Report dateNov 19, 2014
Recall initiation dateSep 30, 2014
Quantity21,754 units (21,222) in the U.S. and Puerto Rico)
LocationIrvine, CA, United States

Official product description

Masimo SET uSpO2 Cable The Masimo SET uSpO2 Pulse Oximetry Cable is an in-line cable that provides continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and pulse rate when installed in a compatible OEM host system. Shipping/Package: All Pulse Oximetry cables are shipped non-sterile in a cardboard carton. They are intended to be used non-sterile.

Why the product was recalled

Firm has identified a small number of Oximetry Cables with crossed-internal wires. The effect on performance when these wires are crossed is that when a low SpO2 value is measured, a high SpO2 value would be displayed on the monitoring device. Additionally when a high SpO2 value is measured, a low SpO2 value would be displayed on the monitoring device.

Distribution information

Worldwide distribution including; nationwide (U.S. and Puerto Rico) and the countries of Argentina, Austria, Brazil, Canada, China, Denmark, France, Gernany, Hong Kong, India, Italy, Japan, Jordan, Kuwait, Mexico, Netherlands, Poland, Portugal, South Korea, Spain, Switzerland, Turkey, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records