FDA recall record

Maquet Getinge Group 1180.19XO Extension Device recall — FDA Z-1045-2016

MAQUET Medical Systems USA is initiating a field action for correction of the MAQUET Extension Device (Part No 1180.19F0). The Extension Device is…

Class IIclassification
Z-1045-2016official record ID
Mar 9, 2016report date

Recall record at a glance

Official recordZ-1045-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMaquet Cardiovascular Us Sales, LLC
Report dateMar 9, 2016
Recall initiation dateJan 7, 2016
Quantity2 within US; 247 OUS
LocationWayne, NJ, United States

Official product description

Maquet Getinge Group 1180.19XO Extension Device. Manufacturer MAQUET GmbH, Used to support the upper and lower extremities immediately before, during and after fracture surgery. The Extension Device (1180.19F0) is a table top module for use with the Maquet MAGNUS operating table system and it is attached to the base mounting point of the MAGNUS operating table.

Why the product was recalled

MAQUET Medical Systems USA is initiating a field action for correction of the MAQUET Extension Device (Part No 1180.19F0). The Extension Device is manufactured by Maquet GmbH, in Germany, and is a table top module for use with the Maquet MAGNUS operating table system. Maquet GmbH was made aware of a potential issue that can result in the Extension Device becoming unlocked from the table top, causing it to fall down.

Distribution information

Worldwide Distribution - US including AZ and MO. Foreign distribution to Australia, Austria, Bahrain, Belgium, Brazil, Canada, Denmark, Ecuador, Estonia, Finland, France, Germany, Hong Kong, India, Italy, Japan, Latvia, Malaysia, Netherlands, Norway, Poland, Portugal, Russia, Saudi Arabia, Serbia, South Korea, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, and Venezuela.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.