Recall record at a glance
| Official record | Z-1630-2018 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Integra LifeSciences Corp. |
| Report date | May 16, 2018 |
| Recall initiation date | Mar 6, 2018 |
| Quantity | 766 cases |
| Location | Plainsboro, NJ, United States |
Official product description
Manuka Honey Wound Gel Catalogue No: 716597 Product Usage: Helps maintain a moist wound environment which has shown to be conducive to wound healing Ideal for minor scalds and burns, minor cuts, lacerations and minor abrasions
Why the product was recalled
Incorrect gel part number listed in a master manufacturing record
Distribution information
US state of RI
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.