FDA recall record

Siemens Medical Solutions USA Mammomat Inspiration system: Product Usage: mammography exams recall — FDA Z-0523-2015

It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the…

Class IIclassification
Z-0523-2015official record ID
Dec 17, 2014report date

Recall record at a glance

Official recordZ-0523-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Medical Solutions USA, Inc.
Report dateDec 17, 2014
Recall initiation dateOct 28, 2014
Quantity3
LocationMalvern, PA, United States

Official product description

Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

Why the product was recalled

It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition Workstation (AWS) table may become unstable and fall over. This may result in a serious injury to the operator.

Distribution information

US Nationwide Distribution in the state of: WV, NY, and FL..

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.