FDA recall record

Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version# recall — FDA Z-0472-2021

Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the…

Class IIclassification
Z-0472-2021official record ID
Nov 25, 2020report date

Recall record at a glance

Official recordZ-0472-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusCompleted
Recalling firmMako Surgical Corporation
Report dateNov 25, 2020
Recall initiation dateOct 23, 2020
Quantity367 units
LocationDavie, FL, United States

Official product description

Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#: THA 3.0 and later, PKA 3.0 and later, and TKA 1.0 - Product Usage: The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO (Mako robot) and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later).

Why the product was recalled

Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan.

Distribution information

Worldwide distribution - US Nationwide distribution and the countries of Australia, Japan, Korea, Netherlands, and Chile, and France, Germany Luxembourg, and UK

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.