Recall record at a glance
| Official record | Z-0472-2021 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Completed |
| Recalling firm | Mako Surgical Corporation |
| Report date | Nov 25, 2020 |
| Recall initiation date | Oct 23, 2020 |
| Quantity | 367 units |
| Location | Davie, FL, United States |
Official product description
Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#: THA 3.0 and later, PKA 3.0 and later, and TKA 1.0 - Product Usage: The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO (Mako robot) and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later).
Why the product was recalled
Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan.
Distribution information
Worldwide distribution - US Nationwide distribution and the countries of Australia, Japan, Korea, Netherlands, and Chile, and France, Germany Luxembourg, and UK
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.