FDA recall record

Siemens Medical Solutions USA MAGNETOM systems, magnetic resonance diagnostic devices (MRDD) recall — FDA Z-0467-2016

Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction,…

Class IIclassification
Z-0467-2016official record ID
Dec 30, 2015report date

Recall record at a glance

Official recordZ-0467-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSiemens Medical Solutions USA, Inc.
Report dateDec 30, 2015
Recall initiation dateDec 7, 2015
Quantityin US 3752 units
LocationMalvern, PA, United States

Official product description

MAGNETOM systems, magnetic resonance diagnostic devices (MRDD).

Why the product was recalled

Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be blown directly into the magnet room or other areas. This could lead to displacement of oxygen, cause cold burns, and/or pose a risk to users, patient or others.

Distribution information

Nationwide including : DC, PR, Virgin Islands and Bermuda

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.