| Official record | Z-1598-2016 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Siemens Medical Solutions USA, Inc. |
| Report date | May 11, 2016 |
| Recall initiation date | Apr 13, 2016 |
| Quantity | 8 systems |
| Location | Malvern, PA, United States |
Official product description
MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.
Why the product was recalled
The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.
Distribution information
US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.