FDA recall record

Randox Laboratories Magnesium on RX Imola analyser IVD recall — FDA Z-0725-2017

According to the firm, Carry over was observed when the amylase or pancreatic amylase test is run directly before or after the Magnesium assay. An update…

Class IIclassification
Z-0725-2017official record ID
Dec 14, 2016report date

Recall record at a glance

Official recordZ-0725-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRandox Laboratories, Limited
Report dateDec 14, 2016
Recall initiation dateApr 14, 2016
Quantity15
LocationCrumlin, N/A, United Kingdom

Official product description

Magnesium on RX Imola analyser IVD

Why the product was recalled

According to the firm, Carry over was observed when the amylase or pancreatic amylase test is run directly before or after the Magnesium assay. An update is required to the RX Imola analyser running order and an acid wash recommended when setting up the magnesium assay. A correction was made to the IFU for the Magnesium assay and all RX Imola customers in the USA were contacted with the updated instruction.

Distribution information

Nationwide Distribution The product was only distributed to US Consignees.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 20, 2019

RANDOX NEFA recall — FDA Z-0367-2020

Non-Esterified Fatty Acids (NEFA) FA115 batch 485343 is not meeting the performance claims and should not be used in any further testing.

Class IIIFDA: TerminatedRandox Laboratories, Limited