FDA recall record

C.R. Bard Magic3 Antibacterial Hydrophilic Intermittent Catheter, Male 14 Fr. (30/Box) recall — FDA Z-0600-2017

Misbranding; the product labeled as an Antibacterial Hydrophilic Intermittent Catheter, is instead an Antibacterial Intermittent Catheter (no Hydrophilic…

Class IIclassification
Z-0600-2017official record ID
Nov 23, 2016report date

Recall record at a glance

Official recordZ-0600-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmC.R. Bard, Inc.
Report dateNov 23, 2016
Recall initiation dateOct 5, 2016
Quantity10,290 units
LocationCovington, GA, United States

Official product description

Magic3 Antibacterial Hydrophilic Intermittent Catheter, Male 14 Fr. (30/Box), Sterile Product Usage: The catheter is an all silicone, single lumen, drainage tube with drainage eyes at the proximal end and a tapered funnel at the distal end. The outer surface of the catheter has an antibacterial and hydrophilic coating which binds water molecules to the surface creating a smooth lubricating film. It is packaged with a water sachet that is manually burst to hydrate the catheter coating at time of use. The catheters are packaged in a film-to-film pouch and are sterilized by electron beam irradiation and are distributed as a single use device. The catheter is intended for urinary bladder drainage in adult males and females requiring catheterization for management of incontinence, voiding dysfunction and surgical procedures. Efficacy of the catheter in preventing urinary tract infection during intermittent use has not been established. The device is not intended to be used as a treatment for active urinary tract infection. For urological use only; urinary catheters are intended for use for bladder management including urine drainage, collection and measurement. The devices are passed to the urinary bladder via the urethra.

Why the product was recalled

Misbranding; the product labeled as an Antibacterial Hydrophilic Intermittent Catheter, is instead an Antibacterial Intermittent Catheter (no Hydrophilic Coating).

Distribution information

US Nationwide Distribution in the states of: CA, FL, IL, IN, MN, MO, NC, NJ, NY, OH, OK, PA, TX and UT.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.