FDA recall record

Hansen Medical Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80 recall — FDA Z-1564-2016

Firm's quarterly dose audit testing showed that one batch was radiated below the augmented dose and therefore did not meet the augmented dose of 28.4 kGy.

Class IIclassification
Z-1564-2016official record ID
May 4, 2016report date

Recall record at a glance

Official recordZ-1564-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHansen Medical Inc.
Report dateMay 4, 2016
Recall initiation dateMar 25, 2016
Quantity8 kits
LocationMountain View, CA, United States

Official product description

Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen Medical Magellan System is a robotic control system that enables positioning and visualization of a Magellan Catheter and an off-the-shelf guide wire while allowing a physician to remain seated and away from the x-ray radiation source. The Accessory Kits contain disposable components that keep the Magellan Catheters and 3rd-party guide wires performing smoothly during system operation.

Why the product was recalled

Firm's quarterly dose audit testing showed that one batch was radiated below the augmented dose and therefore did not meet the augmented dose of 28.4 kGy.

Distribution information

Nationwide Distribution to OH, TX. FL, CA and GA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.