FDA recall record

Fresenius Kabi USA LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004 recall — FDA Z-1282-2024

Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an…

Class IIclassification
Z-1282-2024official record ID
Mar 20, 2024report date

Recall record at a glance

Official recordZ-1282-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusCompleted
Recalling firmFresenius Kabi USA, LLC
Report dateMar 20, 2024
Recall initiation dateSep 16, 2022
Quantity3
LocationNorth Andover, MA, United States

Official product description

LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004

Why the product was recalled

Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.

Distribution information

US Nationwide distribution in the states of WI and NJ.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Other.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records