FDA recall record

GE Healthcare LVivo EF app on Vscan Extend recall — FDA Z-1839-2019

Overestimation bias in automatically calculated ejection fraction (EF) values while using LVivo EF app on the Vscan Extend product.

Class IIclassification
Z-1839-2019official record ID
Jun 26, 2019report date

Recall record at a glance

Official recordZ-1839-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmGE Healthcare, LLC
Report dateJun 26, 2019
Recall initiation dateMay 21, 2019
Quantity119 units
LocationWaukesha, WI, United States

Official product description

LVivo EF app on Vscan Extend. Sold under the following product names: a. Vscan Extend Dual DICOM, Model Numbers H41212ZA, H41212RN; b .Vscan Extend Dual USB, Model Number H41212RK; c. Vscan Extend Dual Wi-Fi, Model Numbers H41212YZ, H41212RL; d. Vscan Extend Sector DICOM, Model Number H41212ZD; e. Vscan Extend Sector USB, Model Number H41212RR; f. Vscan Extend Sector Wi-Fi, Model Number H41212ZC Product Usage: Vscan Extend is a general- purpose diagnostic ultrasound imaging system for use by qualified and trained healthcare professionals enabling visualization and measurement of anatomical structures and fluid. It s pocket-sized portability and simplified user interface enables integration into examination and training sessions indoors and in other environments described in the user manual. The information can be used for basic/focused assessments and adjunctively with other medical data for clinical diagnosis purposes during routine, periodic monitoring, and triage.

Why the product was recalled

Overestimation bias in automatically calculated ejection fraction (EF) values while using LVivo EF app on the Vscan Extend product.

Distribution information

Worldwide distribution - US nationwide distributed in the states of CA, FL, IL, MA , MD, MI, MN, NY, SC, TX, UT. Countries of Australia, China, Croatia, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, Ireland, Italy, Japan, Korea (Republic Of), Netherlands, Romania, Sweden, Switzerland, Thailand, United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records