Recall record at a glance
| Official record | Z-1288-2015 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Ossur H / F |
| Report date | Mar 25, 2015 |
| Recall initiation date | Mar 4, 2015 |
| Quantity | 2,194 (1,464 in U.S. and 730 internationally) |
| Location | Reykjavik, Iceland |
Official product description
LP Rotate Foot System prosthetic device for fitting lower extremity amputations. -Low profile foot module with adjustable shock absorber and rotation with progressive stiffening -Build height: with male pyramid adapter 145 - 153mm (5 11/16 - 6 ) with female pyramid adapter 153 165mm (6 - 6 1/2) -Male pyramid adapter -Female pyramid adapter -Maintenance free, no lubrication required -Maximum weight: 147kg (325lbs) -Available in sizes 22-30
Why the product was recalled
The threads of the outer tube of the shock module of LP Rotate are too short, resulting in insecure assembly. Use of the product may cause the patient to fall with continued use, as the shock module can become loose from the foot blade.
Distribution information
Nationwide within the U.S. and internationally to the following foreign countries; Belgium, United Kingdom, Germany, Czech Republic, France, Israel, Poland, Spain, Australia, Japan, Mexico, Bulgaria, Canada, South Korea, Sweden, Finland,
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.