FDA recall record

Zest Anchors LOCATOR OVERDENTURE IMPLANT SYSTEM STERILE R Rx Only QTY 1 recall — FDA Z-2425-2020

The abutment provided with the Implant System may be out of specification, which could lead to a small gap occurring between the implant and abutment on…

Class IIclassification
Z-2425-2020official record ID
Jul 1, 2020report date

Recall record at a glance

Official recordZ-2425-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmZest Anchors LLC
Report dateJul 1, 2020
Recall initiation dateNov 22, 2019
Quantity154 kits
LocationCarlsbad, CA, United States

Official product description

LOCATOR OVERDENTURE IMPLANT SYSTEM STERILE R Rx Only QTY 1 .

Why the product was recalled

The abutment provided with the Implant System may be out of specification, which could lead to a small gap occurring between the implant and abutment on assembly in the patient.

Distribution information

US Nationwide Distribution in the state of: WI; VA; OH; IL; GA; NC; KS; NJ; CA; WA; OK; MO and Puerto Rico OUS: Switzerland, Germany, Mexico, Poland, and Thailand

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Telephone.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.