FDA recall record

Lithotriptor, Ultrasonic ShockPulse Lithotripsy Transducer recall — FDA Z-0452-2026

The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition,…

Class IIclassification
Z-0452-2026official record ID
Nov 19, 2025report date

Recall record at a glance

Official recordZ-0452-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmOlympus Corporation of the Americas
Report dateNov 19, 2025
Recall initiation dateOct 8, 2025
Quantity1528 units
LocationCenter Valley, PA, United States

Official product description

Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones. Component: ShockPulse Lithotripsy System

Why the product was recalled

The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.

Distribution information

Worldwide distribution - US Nationwide and the countries of Canada, India, Taiwan, Germany, Australia, Singapore.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.