FDA recall record

MicroPort Orthopedics Lineage Shell: REF 36693646, REF 36693848, REF 36694050, REF 36694252 recall — FDA Z-0686-2017

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing…

Class IIIclassification
Z-0686-2017official record ID
Dec 7, 2016report date

Recall record at a glance

Official recordZ-0686-2017
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmMicroPort Orthopedics Inc.
Report dateDec 7, 2016
Recall initiation dateJun 21, 2016
Quantity22 units
LocationArlington, TN, United States

Official product description

Lineage Shell: REF 36693646, REF 36693848, REF 36694050, REF 36694252, REF 36694454, REF 36694656, REF 36694858, REF 36695060, REF 36695262, REF 36695464 For use with the conserve metal head

Why the product was recalled

Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.

Distribution information

Worldwide Distribution - US (nationwide) Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Iran, Italy, Jamaica, Japan, Lithuania, Malaysia, Philippines, Poland, Portugal, Russia, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Netherlands, Turkey, Arab Emirates, United Kingdom, and Vietnam.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.