FDA recall record

Hill-Rom Liko Multirall 200 overhead lift- overhead lift is a general-purpose lift with the recall — FDA Z-0925-2021

Multirall Q-link strap lock has not been fully engaged in the S65 rail carriage hook by user, leading to the Multirall lift becoming detached from the…

Class Iclassification
Z-0925-2021official record ID
Feb 10, 2021report date

Recall record at a glance

Official recordZ-0925-2021
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmHill-Rom, Inc.
Report dateFeb 10, 2021
Recall initiation dateDec 18, 2020
Quantity10470 units Expanded Recall: 154 units
LocationBatesville, IN, United States

Official product description

Liko Multirall 200 overhead lift- overhead lift is a general-purpose lift with the intended use in healthcare, intensive care and rehabilitation. Affected devices: Liko Multirall 200 (product number 3130001) Universal SlingBar 450 R2R (product number 3156095) Universal SlingBar 350 R2R (product number 3156094) Carriage D45 with Double Hook (product number 3136100) Extension belt 300-400 mm (product number 3136226) Extension belt 400-600 mm (product number 3136227) Extension belt 600-1000 mm (product number 3136228) Extension belt 1000-1400mm (product number 3136229)

Why the product was recalled

Multirall Q-link strap lock has not been fully engaged in the S65 rail carriage hook by user, leading to the Multirall lift becoming detached from the rail system, potentially resulting in the fall of both the lift motor and the patient

Distribution information

Worldwide distribution: US (Nationwide) and countries of::AT, AU, BR, CA, CH, CZ, DE, DK, ES, FI, FR, GB, GR, IE, IL, IS, IT, JP, LV, NL, NO, NZ, SE & SK.

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceOct 6, 2021

Hill-Rom LikoStretch Mod 600 IC recall — FDA Z-2601-2021

Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide…

Class IIFDA: OngoingHill-Rom, Inc.
Medical DeviceOct 6, 2021

Hill-Rom LikoStretch 1900 recall — FDA Z-2600-2021

Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide…

Class IIFDA: OngoingHill-Rom, Inc.