Roche Molecular Systems LightMix Zika rRT-PCR Test The LightMix Zika rRT-PCR Test is a recall — FDA Z-0750-2017

The LightCycler 480 algorithm used for the LightMix Zika rRT-PCR Test, EUA (catalog number 07987897001) occasionally (~1% of the time) calls a positive…

Class IIclassification
Z-0750-2017official record ID
Dec 21, 2016report date

Recall record at a glance

Official recordZ-0750-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmRoche Molecular Systems, Inc.
Report dateDec 21, 2016
Recall initiation dateNov 16, 2016
Quantity24 pieces
LocationBranchburg, NJ, United States

Official product description

LightMix Zika rRT-PCR Test The LightMix Zika rRT-PCR Test is a real-time RT-PCR test intended for the qualitative detection of RNA from the Zika virus in serum or EDTA plasma from individuals meeting CDC Zika virus clinical criteria (e.g., clinical signs and symptoms associated with Zika virus infection) and/or CDC Zika virus epidemiological criteria (e.g., history of residence in or travel to a geographic region with active Zika transmission at the time of travel, or other epidemiologic criteria for which Zika virus testing may be indicated). Testing is limited to laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. ¿263a, to perform high complexity tests, or by similarly qualified non-U.S. laboratories. Results are for the identification of Zika virus RNA. Zika virus RNA is generally detectable in serum during the acute phase of infection (approximately 7 days following onset of symptoms, if present). Positive results are indicative of current infection. Laboratories are required to report all positive results to the appropriate public health authorities. Within the United States and its territories positive results must be reported to appropriate state/local public health authorities. Negative results do not preclude Zika virus infection and should not be used as the sole basis for patient management decisions. Negative results must be combined with clinical observations, patient history, and epidemiological information. The LightMix Zika rRT-PCR Test is intended for use by trained clinical laboratory personnel specifically instructed and trained in the techniques of realtime PCR and in vitro diagnostic procedures. The Test is only for use under the Food and Drug Administration s Emergency Use Authorization.

Why the product was recalled

The LightCycler 480 algorithm used for the LightMix Zika rRT-PCR Test, EUA (catalog number 07987897001) occasionally (~1% of the time) calls a positive result for a negative sample if there is an inflection point in the fluorescence signal.

Distribution information

Nationwide Distribution

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.