FDA recall record

Ethicon Endo-Surgery LIGACLIP Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L recall — FDA Z-1777-2013

Ethicon Endo-Surgery is initiating a voluntary recall for LIGACLIP 10mm M/L Endoscopic Rotating Multiple Clip Applier (ER320) due to potential clip…

Class IIclassification
Z-1777-2013official record ID
Jul 31, 2013report date

Recall record at a glance

Official recordZ-1777-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmEthicon Endo-Surgery Inc.
Report dateJul 31, 2013
Recall initiation dateApr 12, 2013
Quantity10,488
LocationCincinnati, OH, United States

Official product description

LIGACLIP Endoscopic Rotating Multiple Clip Applier (ER320), 10mm M/L, bulk non-sterile Product Usage: The LIGACLIP Endoscopic Rotating Multiple Clip Applier is a single patient use instrument designed to provide a means of ligation through ENDOPATH surgical trocars. The instrument delivers 20 titanium clips that individually advance after each firing. The shaft is made of a low glare material that minimizes reflective distortion. It is designed to rotate 360 in either direction. The rotating knob is located to allow for a one-handed technique.

Why the product was recalled

Ethicon Endo-Surgery is initiating a voluntary recall for LIGACLIP 10mm M/L Endoscopic Rotating Multiple Clip Applier (ER320) due to potential clip formation and feeding issues which may result in improper clip formation, vessel damage and/or insufficient occlusion of the vessel or other structure

Distribution information

Worldwide Distribution - USA Nationwide to all fifty (50) states, and internationally including the following countries: ARGENTINA, ARGENTINA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CYPRUS, ECUADOR, EGYPT, EL SALVADOR, GREECE, HONDURAS, INDIA, INDONESIA, ISRAEL, JAMAICA, JAPAN, JORDAN, KOREA, REPUBLIC OF, KUWAIT, LEBANON, MEXICO, NETHERLANDS ANTILLES, NICARAGUA, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, PUERTO RICO, QATAR, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, TAIWAN, PROVINCE OF CHINA, THAILAND, TRINIDAD AND TOBAGO, TURKEY, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA and YEMEN.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.