Hospira LifeShield; Sterile; Rx only; List No. 19666-28; SYMBIQ PUMP SET, CONVERTIBLE PIN recall — FDA Z-1751-2013

Hospira has received reports of customers experiencing air-in-line alarms related to administration sets that have a drip chamber with a ball valve.…

Class IIclassification
Z-1751-2013official record ID
Jul 24, 2013report date

Recall record at a glance

Official recordZ-1751-2013
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHospira Inc.
Report dateJul 24, 2013
Recall initiation dateJun 18, 2013
Quantity14,400 units
LocationLake Forest, IL, United States

Official product description

LifeShield; Sterile; Rx only; List No. 19666-28; SYMBIQ PUMP SET, CONVERTIBLE PIN, PIGGYBACK, 2 CLAVES PORTS AND OPTION LOK, NON-DEHP; NOMINAL LENGTH: 97 IN (249 CM); PRIMING VOLUME: 14 ML; For the administration of I.V. fluids; Made in Costa Rica; Hospira, Inc., Lake Forest, IL 60045 USA The Hospira Symbiq Infusion System is a general infusion system which provides IV infusion across the general spectrum of clinical care. The Symbiq pumps are standard sized pumps which are typically facility-based infusion pumps which may be used to deliver continuous and/or intermittent infusions to a variety of patient populations.

Why the product was recalled

Hospira has received reports of customers experiencing air-in-line alarms related to administration sets that have a drip chamber with a ball valve. Preliminary investigation findings found that there is a potential for the ball valve to not seal the drip chamber as intended, leading to air in the line of the administration set.

Distribution information

Worldwide Distribution - US (nationwide) including the states of IA, IL, MN, SD and WA., and Costa Rica.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 16, 2016

Hospira Symbiq Two Channel Infuser recall — FDA Z-0362-2017

May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The…

Class IIFDA: TerminatedHospira Inc.
Medical DeviceNov 16, 2016

Hospira Symbiq One Channel Infuser recall — FDA Z-0361-2017

May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The…

Class IIFDA: TerminatedHospira Inc.
Medical DeviceNov 9, 2016

Hospira Symbiq Two Channel Infuser recall — FDA Z-0354-2017

Hospira has received reports of Malfunction S205 Backup Battery Failure alarms in Symbiq v3.13 pumps. This issue to an incorrect installation of components on some of the Power Supply…

Class IIFDA: TerminatedHospira Inc.