FDA recall record

Hospira LifeShield LATEX-FREE HemoSet 100 mL Burette I.V recall — FDA Z-1479-2014

Hospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a…

Class IIclassification
Z-1479-2014official record ID
Apr 30, 2014report date

Recall record at a glance

Official recordZ-1479-2014
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHospira Inc.
Report dateApr 30, 2014
Recall initiation dateFeb 25, 2014
Quantity3,893 sets
LocationLake Forest, IL, United States

Official product description

LifeShield LATEX-FREE HemoSet 100 mL Burette I.V. PlumSet, Nonvented, 105 inch with Prepierced Reseals, OPTION-LOK and Inline Dual Channel Cassette. For administration of blood and blood bags for use with Plum Series Infusers.

Why the product was recalled

Hospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a burette and blood filter, intended for infusion of blood/blood products. In a gravity delivery, the correct lower lid dispenses 15 drops per mL and the incorrect lower lid found dispenses 10 drops per mL.

Distribution information

US Distribution including the states of FL, IA,IN, KS, KY, ME, MO, MT, PA, TN, TX, UT, VA and WA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 16, 2016

Hospira Symbiq Two Channel Infuser recall — FDA Z-0362-2017

May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The…

Class IIFDA: TerminatedHospira Inc.
Medical DeviceNov 16, 2016

Hospira Symbiq One Channel Infuser recall — FDA Z-0361-2017

May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The…

Class IIFDA: TerminatedHospira Inc.
Medical DeviceNov 9, 2016

Hospira Symbiq Two Channel Infuser recall — FDA Z-0354-2017

Hospira has received reports of Malfunction S205 Backup Battery Failure alarms in Symbiq v3.13 pumps. This issue to an incorrect installation of components on some of the Power Supply…

Class IIFDA: TerminatedHospira Inc.