Physio-Control LIFEPAK 20e DEFIBRILLATOR/MONITOR, 3201273 Product Usage: The LIFEPAK 20e recall — FDA Z-0294-2018

Physio-Control is aware that some devices have had power-related failures as customers prepared their device for initial deployment or during use within…

Class Iclassification
Z-0294-2018official record ID
Jan 17, 2018report date

Recall record at a glance

Official recordZ-0294-2018
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusTerminated
Recalling firmPhysio-Control, Inc.
Report dateJan 17, 2018
Recall initiation dateDec 4, 2017
Quantity3831 units
LocationRedmond, WA, United States

Official product description

LIFEPAK 20e DEFIBRILLATOR/MONITOR, 3201273 Product Usage: The LIFEPAK 20e defibrillator/monitor is intended for use in the hospital, physician s office, and clinic setting by personnel who are authorized by a physician/medical director. It is intended for use on a crash cart as well as for portable emergency response throughout a hospital.

Why the product was recalled

Physio-Control is aware that some devices have had power-related failures as customers prepared their device for initial deployment or during use within the first year of distribution. The symptoms of these failures may include unexpected power on and power off, device lock-up, or a failure to power on or off, any of which has the potential to result in a failure to deliver therapy to the patient and serious injury or death.

Distribution information

Worldwide Distribution

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.