FDA recall record

DKK Dai-Ichi Shomei Co LFS Flat screen support arm systems without surgical lamp for models recall — FDA Z-0072-2024

The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can…

Class IIclassification
Z-0072-2024official record ID
Oct 11, 2023report date

Recall record at a glance

Official recordZ-0072-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmDKK Dai-Ichi Shomei Co., Ltd.
Report dateOct 11, 2023
Recall initiation dateAug 23, 2023
Quantity1211 units
LocationItabashi, N/A, Japan

Official product description

LFS Flat screen support arm systems without surgical lamp for models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Support Arm Model Numbers

Why the product was recalled

The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.

Distribution information

US States: MI

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.