Recall record at a glance
| Official record | Z-1340-2026 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Elekta, Inc. |
| Report date | Feb 18, 2026 |
| Recall initiation date | Dec 12, 2025 |
| Quantity | N/A |
| Location | Atlanta, GA, United States |
Official product description
Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3, 11.4 Product Description: Leksell GammaPlan is a computer-based system designed for Leksell Gamma Knife treatment planning.
Why the product was recalled
Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.
Distribution information
Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, ST, FL, GA, HI, IA, IL, LA, MA, MI, MN. MO, NC, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Hong Kong, India, Iran, Iraq, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Poland, Romania, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.