Recall record at a glance
| Official record | Z-2112-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Beyond Laser Systems, LLC |
| Report date | Jul 23, 2025 |
| Recall initiation date | Jun 30, 2025 |
| Quantity | 12 |
| Location | Austin, TX, United States |
Official product description
Laser light show projectors manufactured by BLS, Model CLUB 1000/2000/4000.
Why the product was recalled
The remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically installed but not wired. The labels placed on the projectors do not meet requirements specified in 21 CFR 1010.4 (Variances), 21 CFR 1010.2 (Certification), 21 CFR 1010.3 (Identification) and 21 CFR 1040.10 (g)(Labelling requirements). The identification label does not include the date of manufacture as required by 21 CFR 1010.3. The corrective action plan covers three families of laser light show of projectors: DIODE. CLUB and PURE with three different hardware configurations such as the number of the number of RJ45 ports, radiant power of emissions and network configurations. No specific user manuals were provided to each model family as required by 21 CFR 1040.10(h)(1) (Informational requirements (1) User information). The Quality Control test did not validate the manual reset function to enable resumption of laser radiation emission from Class IV laser projectors after an interruption caused using a remote interlock. No supplemental product reports, as required according to 21 CFR 1002.11.
Distribution information
US Nationwide Distribution
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: FDA Mandated. Initial notification: N/A.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.