FDA recall record

Dannoritzer Medizintechnik GmbH & Co. KG Laparoscopy Instruments Instructions For Use (IFU) recall — FDA Z-0062-2019

The Manual cleaning process and the usability of the reprocessing instruction by the user had not been validated. The IFU was modified to remove the…

Class IIIclassification
Z-0062-2019official record ID
Oct 10, 2018report date

Recall record at a glance

Official recordZ-0062-2019
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmDannoritzer Medizintechnik GmbH & Co. KG
Report dateOct 10, 2018
Recall initiation dateAug 6, 2018
LocationTuttlingen, Germany

Official product description

Laparoscopy Instruments Instructions For Use (IFU) for the following: 1)Monopolar electrode, 2) Insert, handle and/or shaft component of a modular forceps or scissors for electrosurgical coagulation, 3) One-piece forceps or scissors for electrosurgical coagulation

Why the product was recalled

The Manual cleaning process and the usability of the reprocessing instruction by the user had not been validated. The IFU was modified to remove the manual cleaning method and revised the instructions for use for clarity and readability.

Distribution information

US Nationwide in the States of OH, MI, MA, CA, OK, TN, GA, FL, IL, NC, TX, MO.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.