Recall record at a glance
| Official record | Z-2197-2019 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Siemens Medical Solutions USA, Inc. |
| Report date | Aug 14, 2019 |
| Recall initiation date | Jul 10, 2019 |
| Quantity | 15 |
| Location | Malvern, PA, United States |
Official product description
Lantis Commander, Model # 4503178 - Product Usage: Lantis Treatstation is to allow the radiation therapist to deliver treatment to the patient using the mevatron and all available accessories. This entails selecting a patient, selecting today s treatment for that patient, setting up and delivering the treatment fields and recording the delivered treatment. Treatstation supports auto sequencing, a process of automatically downloading a group of fields or segments from the verification and record system to the control of the linear accelerator sequentially, without user intervention. In addition, Lantis Treatstation supports intensity modulation, a process of shaping, modifying and moving the beam around a target to maximize the dose at the target and minimize the dose to normal structures.
Why the product was recalled
Potentially affected by the Microsoft Windows Remote Desktop Protocol (RDP) vulnerability
Distribution information
State WV NY PA FL OK FL NJ NY IL WY NY GA NY PA PA AL NJ NY ***updated 8/14/19*** OH NC
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.