Beckman Coulter LabPro Data Management System, UDI: 15099590676872 - Product Usage: The recall — FDA Z-1673-2020

A security-only update was released via remote diagnostics on January 14, 2020 to customers with Windows 7 operation system. Shortly after the releases…

Class IIclassification
Z-1673-2020official record ID
Apr 15, 2020report date

Recall record at a glance

Official recordZ-1673-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBeckman Coulter Inc.
Report dateApr 15, 2020
Recall initiation dateJan 17, 2020
Quantity177 systems
LocationBrea, CA, United States

Official product description

LabPro Data Management System, UDI: 15099590676872 - Product Usage: The LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel.

Why the product was recalled

A security-only update was released via remote diagnostics on January 14, 2020 to customers with Windows 7 operation system. Shortly after the releases some customers reported that Data Management System computers would not restart. If the computer is unable to restart, then the system cannot download data generated from MicroScan instruments or manually entered microbiology test results, resulting in a potential of delayed results.

Distribution information

Worldwide distribution - US Nationwide distribution in the states of Wisconsin, New York, Delaware, Indiana, Ohio, Tennessee, Florida, Georgia, Illinois, South Carolina, California, New Jersey, Alabama, Massachusetts, Michigan, Texas, West Virginia, Wyoming, Missouri, New Mexico, North Carolina, Minnesota, New Hampshire, Oklahoma, Washington, Pennsylvania, Idaho, Mississippi, Maryland, Virginia, Arizona, Utah, Maine, Michigan, Colorado, Iowa, Louisiana, Kentucky, South Dakota, Nevada, North Dakota, Kentucky and countries of Brazil, Germany.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.