FDA recall record

L3 Security & Detection System Cabinet X-ray system used for security screening of checked recall — FDA Z-0270-2018

L3 has determined in airport environments, there have been several cases where a small air gap has developed where the entrance or exit tunnels meet up…

Class IIclassification
Z-0270-2018official record ID
Jan 3, 2018report date

Recall record at a glance

Official recordZ-0270-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmL3 Security & Detection
Report dateJan 3, 2018
Recall initiation dateOct 27, 2017
Quantity471
LocationWoburn, MA, United States

Official product description

L3 Security & Detection System Cabinet X-ray system used for security screening of checked airline baggage.

Why the product was recalled

L3 has determined in airport environments, there have been several cases where a small air gap has developed where the entrance or exit tunnels meet up with the main x-ray scanner tunnel. This a gap in x-ray shielding can result in x-ray emissions that exceed the limit in the federal performance standard for cabinet x-ray systems (see 21 CFR ¿ 1020.40(c)(1)).

Distribution information

USA (nationwide) Distribution.  Owned and operated by the U.S. Transportation Security Administration. Location of individual products is law enforcement sensitive. TSA HQ is located in Crystal City, Virginia.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: FDA Mandated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.