FDA recall record

Halyard Health KimVent Turbo-Cleaning Closed Suction System for Adults, 10 F T-Piece recall — FDA Z-1678-2015

The internal blue flapper valve of certain KimVent Turbo-Cleaning Closed Suction System for Adults may have a potential defect. If present, the defect…

Class IIclassification
Z-1678-2015official record ID
Jun 3, 2015report date

Recall record at a glance

Official recordZ-1678-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmHalyard Health, Inc.
Report dateJun 3, 2015
Recall initiation dateApr 14, 2015
Quantity11 cases each containing 20 individual endotracheal tubes
LocationAlpharetta, GA, United States

Official product description

KimVent Turbo-Cleaning Closed Suction System for Adults, 10 F T-Piece (Product Code 227105); KimVent Turbo-Cleaning Closed Suction System for Adults, 12 F, DSE (Product Code 2271603); KimVent Turbo-Cleaning Closed Suction System for Adults, 12 F, T-Piece (Product Code 227135).

Why the product was recalled

The internal blue flapper valve of certain KimVent Turbo-Cleaning Closed Suction System for Adults may have a potential defect. If present, the defect potentially allows air to leak out of the closed ventilator circuit.

Distribution information

Nationwide Distribution including CA, MA, MI, NC, TX, VA , and WA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.